Calibration FAQs
Signal Generator's 10MHz is used for factory calibration. This applies to all RTSA products.
Wed, Sep 1, 2021 at 12:32 PM
The decimation of 1.
Wed, Sep 1, 2021 at 2:58 PM
Power level of -44dBm is used.
Wed, Sep 1, 2021 at 3:36 PM
For ZIF mode, the signal is injected 20Mhz from the center(tuned frequency). For SH/SHN mode, the input signal is placed at the center. For DD mode, it...
Wed, Sep 1, 2021 at 3:45 PM
Yes, HDR mode is calibrated. However, it is only calibrated when the gain is 25dB.
Tue, Sep 14, 2021 at 10:11 PM